Section 8: TSP command reference

dmm.measure.threshold.autorange

This attribute determines if the threshold range is set manually or automatically.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
state = dmm.measure.threshold.autorange
dmm.measure.threshold.autorange = state
state
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
This command determines how the range is selected.
When this command is set to off, you must set the range. If you do not set the range, the instrument
remains at the range that was selected by autorange.
When this command is set to on, the instrument uses the signal to determine the most sensitive
range on which to perform the measurement. The instrument sets the range when a measurement is
requested. To set the range, the instrument makes a measurement to determine the range before
making the final measurement, which can result in slower reading times. Turn autorange off and set a
specific range to increase measure time.
If a range is manually selected through the front panel or a remote command, this command is
automatically set to off.
Example
dmm.measure.func = dmm.FUNC_ACV_PERIOD
dmm.measure.threshold.autorange = dmm.ON
Also see
dmm.measure.threshold.level
dmm.measure.threshold.range
8-194
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
The auto range setting:
Disable: dmm.OFF
Enable: dmm.ON
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
(on page 8-195)
(on page 8-196)
Default value
dmm.ON
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Set the measure function to period.
Set the threshold autorange on.