[:SENSe[1]]:<function>:COUPling
This command determines if AC or DC signal coupling is used.
Type
Command and query
Usage
[:SENSe[1]]:<function>:COUPling <type>
[:SENSe[1]]:<function>:COUPling?
<function>
<type>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This command selects the type of input coupling that is used for the selected function.
When DC is selected, the instrument measures AC and DC components of the signal. When AC is
selected, the instrument only measures the AC components of the signal.
If AC coupling is selected, you can change input impedance settings, but they do not take effect until
DC coupling is selected.
Example
DIG:VOLT:COUP AC
Also see
DC and AC coupling
[:SENSe[1]]:<function>:COUPling:AC:FILTer
[:SENSe[1]]:<function>:COUPling:AC:FREQuency
[:SENSe[1]]:<function>:INPutimpedance
DMM7510-901-01 Rev. B / May 2015
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The type of coupling:
AC
DC
RESistance
FRESistance
DIODe
CAPacitance
(on page 2-132)
(on page 6-76)
(on page 6-83)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set the digitize volt coupling to AC.
(on page 6-77)
Section 6: SCPI command reference
Default value
DC
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
6-75