Trigger:external:out:stimulus - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference

:TRIGger:EXTernal:OUT:STIMulus

This command selects the event that causes a trigger to be asserted on the external output line.
Type
Command and query
Usage
:TRIGger:EXTernal:OUT:STIMulus <event>
:TRIGger:EXTernal:OUT:STIMulus?
<event>
Details
The trigger stimulus for the external output line can be set to one of the trigger events described in
the following table.
Trigger events
Event description
No trigger event
Front-panel TRIGGER key press
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line <n> (1 to 6)
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
Appropriate LXI trigger packet is received on LAN trigger object
<n> (1 to 8)
Trigger event blender <n> (1 or 2), which combines trigger events BLENder<n>
Trigger timer <n> (1 to 4) expired
Analog trigger
External in trigger
Example
:TRIG:EXT:OUT:STIMulus TIM3
Also see
:TRIGger:EXTernal:OUT:LOGic
6-212
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
The event to use as a stimulus; see Details
(on page 6-211)
Where saved
Save settings
Event constant
NONE
DISPlay
NOTify<n>
COMMand
DIGio<n>
TSPLink<n>
LAN<n>
TIMer<n>
ATRigger
EXTernal
Set the stimulus for the external output to be
the expiration of trigger timer 3.
DMM7510-901-01 Rev. B / May 2015
Default value
NONE