Section 8: TSP command reference

trigger.timer[N].clear()

This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.timer[N].clear()
N
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
trigger.timer[1].clear()
Also see

trigger.timer[N].count

trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event or is enabled as
a timer or alarm.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
count
N
Details
If count is set to a number greater than 1, the timer automatically starts the next trigger timer delay
at the expiration of the previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.
8-328
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Trigger timer number (1 to 4)
(on page 8-328)
Affected by
Restore configuration
Instrument reset
Power cycle
Trigger timer N reset
Number of times to repeat the trigger (0 to 1,048,575)
Trigger timer number (1 to 4)
Where saved
Default value
Where saved
Default value
Configuration script
1
DMM7510-901-01 Rev. B / May 2015
Clears trigger timer 1.