The branch-on-constant-limits block goes to a branching block if a measurement meets the criteria
set by this command.
The type of limit can be:
Above: The measurement is above the value set by limit B; limit A must be set, but is ignored
when this type is selected
Below: The measurement is below the value set by limit A; limit B must be set, but is ignored
when this type is selected
Inside: The measurement is inside the values set by limits A and B; limit A must be the low value
and Limit B must be the high value
Outside: The measurement is outside the values set by limits A and B; limit A must be the low
value and Limit B must be the high value
The measurement block must be a measure or digitize building block that occurs in the trigger model
before the branch-on-constant-limits block. The last measurement from a measure or digitize building
block is used.
If the limit A is more than the limit B, the values are automatically swapped so that the lesser value is
used as the lower limit.
Example
trigger.model.setblock(5, trigger.BLOCK_BRANCH_LIMIT_CONSTANT, trigger.LIMIT_ABOVE,
.1, 1, 2)
Sets trigger block 5 to be a constant limit that branches to block 2 when the measurement is above the value set
for limit B (which is set to 1). Note that limit A must be set, but is ignored.
Also see
Constant Limit block
DMM7510-901-01 Rev. B / May 2015
(on page 3-85)