Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Example
:DIG:LINE3:MODE TRIG, IN
:DIG:LINE5:MODE TRIG, IN
:TRIG:BLEN1:MODE OR
:TRIG:BLEN1:STIM1 DIG3
:TRIG:BLEN1:STIM2 DIG5
Also see
:TRIGger:BLENder<n>:MODE

:TRIGger:BLOCk:BRANch:ALWays

This command defines a trigger model block that always goes to a specific block.
Type
Command only
Usage
:TRIGger:BLOCk:BRANch:ALWays <blockNumber>, <branchToBlock>
<blockNumber>
<branchToBlock>
Details
When the trigger model reaches a branch-always building block, it goes to the building block set by
<branchToBlock>.
Example
TRIG:BLOC:BRAN:ALW 9, 20
Also see
None
DMM7510-901-01 Rev. B / May 2015
(on page 6-178)
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
The block number of the trigger model block to execute when the trigger model
reaches this block
Section 6: SCPI command reference
Set digital I/O lines 3 and 5 as trigger in
lines. Generate a trigger blender 1 event
when a digital I/O trigger happens on line 3
or 5.
Where saved
Save settings
When the trigger model reaches block 9, it
will always branch to block 20.