Section 8: TSP command reference

trigger.tsplinkout[N].logic

This attribute defines the trigger output with output logic for a trigger line.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
logicType = trigger.tsplinkout[N].logic
trigger.tsplinkout[N].logic = logicType
logicType
N
Details
This attribute controls the logic that the output trigger generator uses on the given trigger line.
The output state of the digital I/O line is controlled by the trigger logic, and the user-specified output
state of the line is ignored.
Example
tsplink.line[3].mode = tsplink.MODE_TRIGGER_OPEN_DRAIN
trigger.tsplinkout[3].logic = trigger.LOGIC_POSITIVE
Also see
trigger.tsplinkout[N].assert()
tsplink.line[N].mode

trigger.tsplinkout[N].pulsewidth

This attribute sets the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
width = trigger.tsplinkout[N].pulsewidth
trigger.tsplinkout[N].pulsewidth = width
width
N
8-342
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
TSP-Link line N reset
The output logic of the trigger generator:
Assert a TTL-high pulse for output: trigger.LOGIC_POSITIVE
Assert a TTL-low pulse for output: trigger.LOGIC_NEGATIVE
The trigger line (1 to 3)
(on page 8-341)
(on page 8-348)
Affected by
Restore configuration
Instrument reset
Power cycle
TSP-Link line N reset
The pulse width (0.0 to 100 ks)
The trigger line (1 to 3)
Where saved
Default value
Configuration script
trigger.LOGIC_NEGATIVE
Sets the trigger logic for
synchronization line 3 to output a
positive pulse.
Where saved
Default value
10e-6 (10 µs)