Section 2: General operation
Setting
Show Warning
Show Information
Popups
Log Warning
Log Information
Log Command
Reset Popups
Save to USB
Clear Log
2-50
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Description
Turns the display of warnings on or off. If you turn this off, the
instrument continues to record warnings and display warning
popup messages, but does not display them on the System Events
tab.
Turns the display of information messages on or off. If you turn this
off, the instrument continues to record information messages and
display popup messages, but does not display them on the System
Events tab.
Chooses what type and whether to display popup messages on the
front panel. You can choose to display error messages, error and
warning messages, or no messages in popups. Messages continue
to be saved in the event log when popups are turned off.
Turns the logging of warnings on or off. If this is turned off, the
instrument does not log or display popups for warning messages.
Turns the logging of information messages on or off. If this is
turned off, the instrument does not log or display popups for
information messages.
Turns the logging of commands on or off. When logging is turned
on, the instrument records the commands that are sent to the
instrument. It records commands sent from any interface (the front
panel or a remote interface).
Restores the popups setting to show errors and warnings.
Saves the event log to a .csv file on the USB flash drive. The
filename is eventlog.csv.
Clears all entries from the event log.