Keithley DMM7510 Reference Manual page 939

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference
All items in the list are subject to the same action; you cannot combine AND and OR logic in a single
block.
You cannot leave the first event as no trigger. If the first event is not defined, the trigger model errors
when you attempt to initiate it.
The following table shows the constants for the events.
Trigger events
Event description
No trigger event
Front-panel TRIGGER key press
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
(1 to 6)
Line edge detected on TSP-Link synchronization line N
(1 to 3)
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
Analog trigger
External in trigger
Example
trigger.model.setblock(9, trigger.BLOCK_WAIT, trigger.EVENT_DISPLAY)
Set trigger model block 9 to wait for a user to press the TRIGGER key on the front panel before continuing.
Also see
Wait block
(on page 3-76)
8-326
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Event constant
trigger.EVENT_NONE
trigger.EVENT_DISPLAY
trigger.EVENT_NOTIFYN
trigger.EVENT_COMMAND
trigger.EVENT_DIGION
trigger.EVENT_TSPLINKN
trigger.EVENT_LANN
trigger.EVENT_BLENDERN
trigger.EVENT_TIMERN
trigger.EVENT_ANALOGTRIGGER
trigger.EVENT_EXTERNAL
DMM7510-901-01 Rev. B / May 2015