System:posetup - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference

:SYSTem:POSetup

This command selects the defaults that are used when you power on the instrument.
Type
Command and query
Usage
:SYSTem:POSetup <name>
:SYSTem:POSetup?
<name>
Details
When you select RST, the instrument restores settings to their default values when the instrument is
powered on.
When you select a SAV option, the settings in the selected saved setup are applied when the
instrument is powered on. The settings are saved using the *SAV command.
Example
SYST:POS SAV1
Also see
*SAV
(on page 6-14)
6-148
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Not applicable
Which setup to restore when you power on the instrument:
Power on to *RST defaults: RST
Stored setup 0: SAV0
Stored setup 1: SAV1
Stored setup 2: SAV2
Stored setup 3: SAV3
Stored setup 4: SAV4
Where saved
Nonvolatile memory
Set the instrument to restore the settings
that are saved in the stored setup 1 when
the instrument is powered on.
DMM7510-901-01 Rev. B / May 2015
Default value
RST

Advertisement

Table of Contents
loading

Table of Contents