Trigger.blender[N].Overrun - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = trigger.EVENT_DIGIO3
trigger.blender[1].stimulus[2] = trigger.EVENT_DIGIO5
Also see
trigger.blender[N].reset()

trigger.blender[N].overrun

This attribute indicates whether or not an event was ignored because of the event detector state.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
overrun = trigger.blender[N].overrun
overrun
N
Details
Indicates if an event was ignored because the event detector was already in the detected state when
the event occurred. This is an indication of the state of the event detector that is built into the event
blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any
other trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
Also see
trigger.blender[N].reset()
DMM7510-901-01 Rev. B / May 2015
(on page 8-256)
Affected by
Where saved
Instrument reset
Not applicable
Trigger blender N clear
Trigger blender N reset
Trigger blender overrun state (true or false)
The blender number (1 or 2)
(on page 8-256)
Section 8: TSP command reference
Generate a trigger blender 1
event when a digital I/O trigger
happens on line 3 or 5.
Default value
Not applicable
If an event was ignored, the output
is true.
If an event was not ignored, the
output is false.
8-255

Advertisement

Table of Contents
loading