[:SENSe[1]]:<function>:COUPling:AC:FILTer
This command selects the instrument settling time when coupling is set to AC.
Type
Command and query
Usage
[:SENSe[1]]:<function>:COUPling:AC:FILTer <type>
[:SENSe[1]]:<function>:COUPling:AC:FILTer?
<function>
<type>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This option is only used when digitize signal coupling is set to AC.
When the signal coupling is set to AC, there may still be some DC signal content that comes in with
the AC signal. To allow this signal to settle out, you can set AC coupling filter to slow. When the filter
is set to slow, the instrument adds an 800 ms delay before making measurements.
When the AC coupling filter is set to fast, the instrument adds an 80 ms delay before making
measurements. Set the AC filter to fast for faster settling when measuring rapidly changing inputs.
For most digitize voltage measurements, the 80 ms delay is enough time for the range to settle.
Example
DIG:VOLT:COUP AC
DIG:VOLT:COUP:AC:FILT FAST
Also see
DC and AC coupling
[:SENSe[1]]:<function>:COUPling
6-76
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
Type of AC coupling filter:
Slow (800 ms): SLOW
Fast (80 ms): FAST
RESistance
FRESistance
DIODe
CAPacitance
(on page 2-132)
(on page 6-75)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set the coupling type to AC.
Set the filter to fast.
DMM7510-901-01 Rev. B / May 2015
Default value
SLOW
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent