Calculate[1]:<Function>:Math:format - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This command sets the high limit for the limit Y test for the selected measurement function. When limit
Y testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Default is 0.8 for limit 1 when the diode function is selected; 10 when the continuity function is
selected. The default for limit 2 for the diode and continuity functions is 1.
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Also see
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
:CALCulate2:<function>:LIMit<Y>:STATe
:CALCulate[1]:<function>:MATH:FORMat
This command specifies which math operation is performed on measurements when math operations are
enabled.
Type
Command and query
Usage
:CALCulate[1]:<function>:MATH:FORMat <operation>
:CALCulate[1]:<function>:MATH:FORMat?
<function>
<operation>
DMM7510-901-01 Rev. B / May 2015
RESistance
FRESistance
DIODe
CAPacitance
(on page 6-29)
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The name of the math operation:
y = mx+b: MXB
Percent: PERCent
Reciprocal: RECiprocal
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage
measurement exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results
display on the front panel
Return the test results; example output if the test
fails on the low limit:
LOW
Clear the test results.
(on page 6-27)
Where saved
Save settings
Measure configuration list
Section 6: SCPI command reference
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
Default value
PERC
6-31