[:Sense[1]]:<Function>:Atrigger:edge:level - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel
This command defines the signal level that generates the analog trigger event for the edge trigger mode.
Type
Command and query
Usage
[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel <setting>
[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel?
<function>
<setting>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This command is only available when the analog trigger mode is set to edge.
The edge level can be set to any value in the active measurement range. See the Model DMM7510
specifications for more information on the resolution and accuracy of the analog trigger.
To use the analog trigger with the measure functions, a range must be set (you cannot use
autorange) and autozero must be disabled.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE EDGE
CURR:ATR:EDGE:LEV 2.5
Also see
Analog triggering overview
[:SENSe[1]]:<function>:ATRigger:MODE
[:SENSe[1]]:<function>:AZERo[:STATe]
[:SENSe[1]]:<function>:SENSe:RANGe[:UPPer]
6-56
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The signal level that generates the trigger event
RESistance
FRESistance
DIODe
CAPacitance
(on page 3-64)
(on page 6-59)
(on page 6-72)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to edge.
Set the analog trigger level to 2.5 A.
(on page 6-105)
DMM7510-901-01 Rev. B / May 2015
Default value
0
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent

Advertisement