Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Details
The trigger stimulus for the external output line can be set to one of the trigger events described in
the following table.
Trigger events
Event description
No trigger event
Front-panel TRIGGER key press
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
(1 to 6)
Line edge detected on TSP-Link synchronization line N
(1 to 3)
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
Analog trigger
External in trigger
Example
trigger.extout.stimulus = trigger.EVENT_TIMER3
Also see
trigger.extin.edge
trigger.extin.wait()
trigger.extout.assert()
trigger.extout.logic
DMM7510-901-01 Rev. B / May 2015
(on page 8-269)
(on page 8-270)
(on page 8-271)
(on page 8-271)
Section 8: TSP command reference
Event constant
trigger.EVENT_NONE
trigger.EVENT_DISPLAY
trigger.EVENT_NOTIFYN
trigger.EVENT_COMMAND
trigger.EVENT_DIGION
trigger.EVENT_TSPLINKN
trigger.EVENT_LANN
trigger.EVENT_BLENDERN
trigger.EVENT_TIMERN
trigger.EVENT_ANALOGTRIGGER
trigger.EVENT_EXTERNAL
Set the stimulus for the external output to
be the expiration of trigger timer 3.
8-273