Dmm.measure.limit[Y].High.value - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

dmm.measure.limit[Y].high.value

This attribute specifies the upper limit for a limit test.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
highLimit = dmm.measure.limit[Y].high.value
dmm.measure.limit[Y].high.value = highLimit
highLimit
Y
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
This command sets the high limit for the limit Y test for the selected measurement function. When limit
Y testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Default is 0.8 for limit 1 when the diode function is selected; 10 when the continuity function is
selected. The default for limit 2 for the diode and continuity functions is 1.
Example
See
dmm.measure.limit[Y].low.value
Also see
dmm.digitize.limit[Y].high.value
dmm.measure.limit[Y].enable
dmm.measure.limit[Y].low.value
DMM7510-901-01 Rev. B / May 2015
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
The value of the upper limit (-1e+12 to 1e+12)
Limit number: 1 or 2
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
(on page 8-164) for an example of how to use this command.
(on page 8-99)
(on page 8-161)
(on page 8-164)
Section 8: TSP command reference
Default value
1 for most functions; see Details
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
8-163