[:SENSe[1]]:<function>:ATRigger:PULSe:WIDTh
This command defines the threshold value for the pulse width.
Type
Command and query
Usage
[:SENSe[1]]:<function>:ATRigger:PULSe:WIDTh <setting>
[:SENSe[1]]:<function>:ATRigger:PULSe:WIDTh?
<function>
<setting>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This option is only available when the analog trigger mode is set to pulse.
This option sets either the minimum or maximum pulse width that generates an analog trigger event.
The value of pulse condition determines whether this value is interpreted as the minimum or
maximum pulse width.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE PULSE
CURR:ATR:PULS:LEV 2.5
CURR:ATR:PULS:POL BEL
CURR:ATR:PULS:WIDT 30e-6
Also see
Analog triggering overview
[:SENSe[1]]:<function>:ATRigger:MODE
[:SENSe[1]]:<function>:ATRigger:PULSe:CONDition
[:SENSe[1]]:<function>:ATRigger:PULSe:LEVel
DMM7510-901-01 Rev. B / May 2015
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The threshold value for the pulse width: 1 µs to 40 ms
RESistance
FRESistance
DIODe
CAPacitance
(on page 3-64)
(on page 6-59)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to pulse.
Set the analog trigger level to 2.5 A.
Set the polarity to below.
Set the analog trigger pulse width to 30 µs.
(on page 6-60)
(on page 6-61)
Section 6: SCPI command reference
Default value
20 µs
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
6-63