Section 8: TSP command reference

tsplink.line[N].mode

This attribute defines the trigger operation of a TSP-Link line.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
mode = tsplink.line[N].mode
tsplink.line[N].mode = mode
mode
N
Details
This command defines whether or not the line is used as a digital or trigger control line and if it is an
input or output.
The line mode can be set to the following options:
TSP-Link digital open drain line: tsplink.MODE_DIGITAL_OPEN_DRAIN
TSP-Link trigger open drain line: tsplink.MODE_TRIGGER_OPEN_DRAIN
TSP-Link trigger synchronous master: tsplink.MODE_SYNCHRONOUS_MASTER
TSP-Link trigger synchronous acceptor: tsplink.MODE_SYNCHRONOUS_ACCEPTOR
Example
tsplink.line[3].mode = tsplink.MODE_TRIGGER_OPEN_DRAIN
Also see
trigger.tsplinkin[N].edge
trigger.tsplinkout[N].logic
8-348
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
TSP-Link line N reset
The trigger mode; see Details
The trigger line (1 to 3)
(on page 8-339)
(on page 8-342)
Where saved
Default value
Configuration script
tsplink.MODE_DIGITAL_OPEN_DRAIN
Sets the trigger mode for
synchronization line 3 as a
trigger open drain line.