Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Example
dmm.measure.func = dmm.FUNC_DC_VOLTAGE
dmm.measure.range = 10
dmm.measure.autozero.enable = dmm.OFF
dmm.measure.analogtrigger.mode = dmm.MODE_PULSE
dmm.measure.analogtrigger.pulse.level = 5
dmm.measure.analogtrigger.pulse.width = 30e-6
dmm.measure.analogtrigger.pulse.condition = dmm.CONDITION_LESS
dmm.measure.analogtrigger.pulse.polarity = dmm.POLARITY_BELOW
Set measure function to DC voltage.
Set range to 10 V.
Disable autozero.
Set the analog trigger mode to pulse.
Set the analog trigger level to 5 V.
Set the analog trigger pulse width to 30 µs.
Set the condition to be detect trigger within the pulse width.
Set the trigger to occur when the pulse is below the level.
Also see
Analog triggering overview
dmm.digitize.analogtrigger.pulse.width
dmm.measure.analogtrigger.mode
dmm.measure.analogtrigger.pulse.condition
dmm.measure.analogtrigger.pulse.polarity
dmm.measure.analogtrigger.pulse.width
DMM7510-901-01 Rev. B / May 2015
(on page 3-64)
(on page 8-79)
(on page 8-121)
(on page 8-122)
(on page 8-125)
(on page 8-126)