Trace:actual:end - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
Example
TRACe:MAKE "testData", 200
COUN 10
MEASure:CURRent? "testData"
:TRACe:ACTual?
:TRACe:ACTual? "testData"
Also see
Reading buffers
Remote buffer operation
:TRACe:MAKE

:TRACe:ACTual:END?

This command indicates the last index in a reading buffer.
Type
Query only
Usage
:TRACe:ACTual:END?
:TRACe:ACTual:END? "<bufferName>"
<bufferName>
Details
Use this command to find the ending index in a reading buffer.
6-152
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 3-13)
(on page 3-30)
(on page 6-160)
Affected by
Recall settings
Instrument reset
Power cycle
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Creates 200 element reading buffer named testData.
Set the measurement count to 10.
Set the measurement function to current. Make readings, and
store the readings in testData. Returns the 10
measurement reading after taking all 10 readings.
Returns the number of readings in defbuffer1.
Example output:
850
Returns the number of readings in the buffer testData.
Example output:
10
Where saved
Not applicable
th
Default value
Not applicable
DMM7510-901-01 Rev. B / May 2015