Tspnet.reset(); Tspnet.termination() - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

tspnet.reset()

This function disconnects all TSP-Net sessions.
Type
TSP-Link accessible
Function
No
Usage
tspnet.reset()
Details
This command disconnects all remote instruments connected through TSP-Net. For TSP-enabled
devices, this causes any commands or scripts running remotely to be terminated.
Also see
None

tspnet.termination()

This function sets the device line termination sequence.
Type
TSP-Link accessible
Function
No
Usage
type = tspnet.termination(connectionID)
type = tspnet.termination(connectionID, termSequence)
type
connectionID
termSequence
Details
This function sets and gets the termination character sequence that is used to indicate the end of a
line for a TSP-Net connection.
Using the termSequence parameter sets the termination sequence. The present termination
sequence is always returned.
For the termSequence parameter, use the same values listed in the table above for type. There are
four possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns
(CR or 0x13). For TSP-enabled devices, the default is tspnet.TERM_LF. For devices that are not
TSP-enabled, the default is tspnet.TERM_CRLF.
DMM7510-901-01 Rev. B / May 2015
Affected by
Where saved
Affected by
Where saved
The termination type:
tspnet.TERM_LF
tspnet.TERM_CR
tspnet.TERM_CRLF
tspnet.TERM_LFCR
The connection ID returned from tspnet.connect()
The termination sequence:
tspnet.TERM_LF
tspnet.TERM_CR
tspnet.TERM_CRLF
tspnet.TERM_LFCR
Section 8: TSP command reference
Default value
Default value
8-359

Advertisement

Table of Contents
loading

Table of Contents