[:Sense[1]]:<Function>:Rtd:beta - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
[:SENSe[1]]:<function>:RTD:BETA
This command contains the beta value of a user-defined RTD.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RTD:BETA <value>
[:SENSe[1]]:<function>:RTD:BETA DEFault
[:SENSe[1]]:<function>:RTD:BETA MINimum
[:SENSe[1]]:<function>:RTD:BETA MAXimum
[:SENSe[1]]:<function>:RTD:BETA?
[:SENSe[1]]:<function>:RTD:BETA? DEFault
[:SENSe[1]]:<function>:RTD:BETA? MINimum
[:SENSe[1]]:<function>:RTD:BETA? MAXimum
<function>
<value>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
This attribute is only valid when:
The function is set to temperature.
The transducer type is set to 3 or 4-wire RTD.
The RTD type is set to user-defined.
Example
:FUNC "TEMP"
:TEMP:TRANsducer TRTD
:TEMP:RTD:THR USER
:TEMP:RTD:ALPH .005
:TEMP:RTD:DELT .00385
:TEMP:RTD:ZERO 120
:TEMP:RTD:BETA .3
Also see
[:SENSe[1]]:<function>:RTD:FOUR
[:SENSe[1]]:<function>:RTD:THRee
[:SENSe[1]]:<function>:TRANsducer
6-98
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
0 to 1
RESistance
FRESistance
DIODe
CAPacitance
(on page 6-100)
(on page 6-101)
(on page 6-112)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set the measure function to temperature.
Set the transducer type to 3-wire RTD.
Set the RTD type to User.
Set the alpha RTD value to 0.005.
Set the delta RTD value to 0.00385.
Set the zero RTD value to 120.
Set the beta RTD value to 0.3.
DMM7510-901-01 Rev. B / May 2015
Default value
0.10863
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent

Advertisement