Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

:TRIGger:LOAD "SimpleLoop"

This command loads a predefined trigger model configuration.
Type
Command only
Usage
:TRIGger:LOAD "SimpleLoop", <count>
:TRIGger:LOAD "SimpleLoop", <count>, <delay>
:TRIGger:LOAD "SimpleLoop", <count>, <delay>, "<bufferName>"
:TRIGger:LOAD "SimpleLoop", <count>, <delay>, "<bufferName>", <readingBlock>
<count>
<delay>
<bufferName>
<readingBlock>
Details
This command sets up a loop that sets a delay, makes a measurement, and then repeats the loop the
number of times you define in the count parameter.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen.
You can use the TRIGger:BLOCk:LIST? command to view the trigger model blocks in a list format.
Example
*RST
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
TRIG:LOAD "SimpleLoop", 10
INIT
*WAI
TRAC:DATA? 1, 10, "defbuffer1", READ, REL
Also see
None
DMM7510-901-01 Rev. B / May 2015
Affected by
Not applicable
The number of measurements the instrument will make
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
A string that indicates the reading buffer; the default buffers (defbuffer1
or defbuffer2) or the name of a user-defined buffer; if no buffer is
specified, defbuffer1 is used
Define a measure or digitize block for the trigger model; options are:
ACTive: Add a measure or digitize block to the trigger model based
on the active function; if no option defined, ACTive is used
MEASure: Adds a measure block to the trigger model
DIGitize: Adds a digitize block to the trigger model
Section 6: SCPI command reference
Where saved
Not applicable
Reset the instrument and set it to
measure current with automatic range
setting.
Set a simple trigger loop with a count of
10.
Start the trigger model.
Postpone execution of subsequent
commands until all previous commands
are finished.
Read data and return the reading and
relative time.
Default value
Not applicable
6-231