Trigger.ext.reset() - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Trigger events
Event description
No trigger event
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
Analog trigger
External in trigger
Example
digio.line[2].mode = digio.MODE_TRIGGER_OUT
trigger.digout[2].stimulus = trigger.EVENT_TIMER3
Also see
digio.line[N].mode
digio.line[N].reset()
Digital I/O port configuration
trigger.digin[N].clear()
trigger.digout[N].assert()

trigger.ext.reset()

This function resets the edge, logic, and stimulus values for the external in/out port to their default values.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.ext.reset()
Details
This function resets the following attributes to their default values:
trigger.extin.edge
trigger.extout.logic
trigger.extout.stimulus
It also clears trigger.extin.overrun.
DMM7510-901-01 Rev. B / May 2015
(on page 8-52)
(on page 8-54)
(on page 3-49)
(on page 8-260)
(on page 8-263)
Affected by
Section 8: TSP command reference
Event constant
trigger.EVENT_NONE
trigger.EVENT_BLENDERN
trigger.EVENT_TIMERN
trigger.EVENT_ANALOGTRIGGER
trigger.EVENT_EXTERNAL
Set the stimulus for output digital
trigger line 2 to be the expiration of
trigger timer 3.
Where saved
Default value
8-267