[:SENSe[1]]:<function>:ATRigger:PULSe:POLarity
This command defines the polarity of the pulse that generates an analog trigger event.
Type
Command and query
Usage
[:SENSe[1]]:<function>:ATRigger:PULSe:POLarity <setting>
[:SENSe[1]]:<function>:ATRigger:PULSe:POLarity?
<function>
<setting>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
Only used when analog trigger mode is pulse.
Determines if the analog trigger occurs when the pulse is above the defined signal level or below the
defined signal level.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE PULSE
CURR:ATR:PULS:LEV 2.5
CURR:ATR:PULS:POL BEL
Also see
Analog triggering overview
[:SENSe[1]]:<function>:ATRigger:PULSe:LEVel
[:SENSe[1]]:<function>:ATRigger:MODE
6-62
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The setting:
Above: ABOVe
Below: BELow
RESistance
FRESistance
DIODe
CAPacitance
(on page 3-64)
(on page 6-59)
Where saved
Configuration script
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to pulse.
Set the analog trigger level to 2.5 A.
Set the polarity to below.
(on page 6-61)
Default value
ABOVe
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
DMM7510-901-01 Rev. B / May 2015