dmm.measure.autozero.once()

This function causes the instrument to refresh the reference and zero measurements once.
Type
TSP-Link accessible
Function
Yes
Usage
dmm.measure.autozero.once()
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
This command forces a refresh of the reference and zero measurements that are used for the present
aperture setting for the selected function.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the
drift, you can send the once command to make a reference and zero measurement immediately
before a test sequence.
If the NPLC setting is less than 0.2 PLC, sending autozero once can result in delay of more than a
second.
Example
dmm.measure.autozero.once()
Also see
Automatic reference measurements
dmm.measure.autozero.enable
8-136
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Where saved
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
(on page 2-149)
(on page 8-135)
Default value
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Do a one-time refresh of the reference and zero
measurements.
DMM7510-901-01 Rev. B / May 2015

Advertisement

Table of Contents
loading

Table of Contents