Dmm.measure.math.format - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference

dmm.measure.math.format

This attribute specifies which math operation is performed on measurements when math operations are enabled.
Type
TSP-Link accessible
Attribute (RW) Yes
Usage
operation = dmm.measure.math.format
dmm.measure.math.format = operation
operation
Details
This specifies which math operation is performed on measurements for the selected measurement
function.
You can choose one of the following math operations:
y = mx+b: Manipulate normal display readings by adjusting the m and b factors.
Percent: Displays measurements as the percentage of deviation from a specified reference constant.
Reciprocal: The reciprocal math operation displays measurement values as reciprocals. The displayed
value is 1/X, where X is the measurement value (if relative offset is being used, this is the measured
value with relative offset applied).
Math calculations are applied to the input signal after relative offset and before limit tests.
Example
dmm.measure.func = dmm.FUNC_DC_VOLTAGE
dmm.measure.math.format = dmm.MATH_RECIPROCAL
dmm.measure.math.enable = dmm.ON
Also see
Calculations that you can apply to measurements
dmm.digitize.math.format
dmm.measure.math.enable
8-168
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
Math operation to be performed on measurements:
y = mx+b: dmm.MATH_MXB
Percent: dmm.MATH_PERCENT
Reciprocal: dmm.MATH_RECIPROCAL
(on page 3-7)
(on page 8-103)
(on page 8-167)
Default value
dmm.MATH_PERCENT
Enables the reciprocal math operation on
voltage measurements.
DMM7510-901-01 Rev. B / May 2015

Advertisement

Table of Contents
loading

Table of Contents