System:eventlog:post - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Example
SYST:EVEN:NEXT?
Also see
:SYSTem:CLEar
:SYSTem:EVENtlog:SAVE

:SYSTem:EVENtlog:POST

This command allows you to post your own text to the event log.
Type
Command only
Usage
:SYSTem:EVENtlog:POST "<message>"
:SYSTem:EVENtlog:POST "<message>", <eventType>
<message>
<eventType>
Details
You can use this command to create your own event log entries and assign a severity level to them.
This can be useful for debugging and status reporting.
From the front panel, you must set the Log Warnings and Log Information options on to have the
custom warning and information events placed into the event log.
Example
*CLS
SYST:EVEN:POST "my error", INF
SYST:EVEN:NEXT?
Also see
None
DMM7510-901-01 Rev. B / May 2015
Returns information on the next event in the event log. For example, if you
sent a command without a parameter, the return is:
-109,"Missing parameter;1;2015/05/06 12:55:33.648"
(on page 6-136)
(on page 6-144)
Affected by
Not applicable
A string that contains the message that will be associated with this event
The type of event that is generated; set to:
The error type: ERRor
The warning type: WARNing
The informational type: INFormational (default)
Section 6: SCPI command reference
Where saved
Not applicable
Clear the event log.
Post an error named my error.
Output:
1003,"User: my error;4,1400469179,431599191"
Default value
Not applicable
6-143