Section 6: SCPI command reference

:TRIGger:BLOCk:LIST?

This command returns the settings for all trigger model blocks.
Type
Query only
Usage
:TRIGger:BLOCk:LIST?
Details
This returns the settings for the trigger model.
Example
:TRIG:BLOC:LIST?
Also see
None

:TRIGger:BLOCk:LOG:EVENt

This command allows you to log an event in the event log when the trigger model is running.
Type
Command only
Usage
:TRIGger:BLOCk:LOG:EVENt <blockNumber>, <eventNumber>, "<message>"
<blockNumber>
<eventNumber>
<message>
6-198
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Not applicable
Returns the settings for the trigger model. Example output is:
1) BUFFER_CLEAR
2) MEASURE
3) BRANCH_COUNTER
4) DELAY_CONSTANT
5) BRANCH_COUNTER
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
The event number:
INFO<n>
WARNing<n>
ERRor<n>
Where <n> is 1 to 4; you can define up to four of each type
You can also set ABORt, which aborts the trigger model immediately and posts a
warning event log message
A string up to 31 characters
Where saved
Not applicable
BUFFER: defbuffer1
BUFFER: defbuffer1 COUNT: 1
VALUE: 5
BRANCH_BLOCK: 2
DELAY: 1.000000000
VALUE: 3
BRANCH_BLOCK: 2
Where saved
Save settings
DMM7510-901-01 Rev. B / May 2015
Default value
Not applicable
Default value
Not applicable