This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
:TRIG:TIM1:CLEar
Also see
:TRIGger:TIMer<n>:COUNt
:TRIGger:TIMer<n>:STARt:OVERrun?
:TRIGger:TIMer<n>:COUNt
This command sets the number of events to generate each time the timer generates a trigger event or is enabled
as a timer or alarm.
Type
Command and query
Usage
:TRIGger:TIMer<n>:COUNt <count>
:TRIGger:TIMer<n>:COUNt?
<n>
<count>
Details
If count is set to a number greater than 1, the timer automatically starts the next trigger timer delay
at the expiration of the previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.
Example 1
TRIG:TIM2:COUN 4
DMM7510-901-01 Rev. B / May 2015
(on page 6-235)
(on page 6-239)
Affected by
Recall settings
Instrument reset
Power cycle
Trigger timer number (1 to 4)
The number of times to repeat the trigger (0 to 1,048,575)
Section 6: SCPI command reference
Clears trigger timer 1.
Where saved
Save settings
Set the number of events to generate for trigger
timer 2 to four.
Default value
1