Reviewing Calibration Information - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 3: Functions and features
From the front panel:
1. Press the MENU key.
2. Under System, select Calibration.
3. Select Scheduling Action. To have the instrument:
Prompt you to run auto calibration: Select Notify.
Run auto calibration at a specific time: Select Run.
To stop scheduling: Select None. If you select None, you do not need to make additional
settings.
4. Select Scheduling Interval.
5. Select the interval.
6. Select Scheduled Time to select the time when the auto calibration will run or when you will be
prompted to run it.
To review the next schedule time and date, see the information listed next to Next Run.
Using SCPI commands:
Refer to
:ACAL:SCHedule
Using TSP commands:
Refer to
acal.schedule()

Reviewing calibration information

The Calibration screen displays information about the last autocalibration and factory calibrations that
were run and the present status. For detail on this screen, refer to
2-53).
For autocalibration, you can also access this information from the commands in the SCPI
subsystem
(on page 6-15) or the TSP acal.* commands.
3-46
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 6-21).
(on page 8-13).
System Calibration menu
ACAL
DMM7510-901-01 Rev. B / May 2015