Keithley DMM7510 Reference Manual page 923

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference
The following table shows the constants for the events.
Trigger events
Event description
No trigger event
Front-panel TRIGGER key press
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
(1 to 6)
Line edge detected on TSP-Link synchronization line N
(1 to 3)
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
Analog trigger
External in trigger
Example
trigger.model.setblock(6, trigger.BLOCK_BRANCH_ON_EVENT, trigger.EVENT_DISPLAY,
2)
When the trigger model reaches this block, if the front-panel TRIGGER key has been pressed, the trigger
model returns to block 2. If the TRIGGER key has not been pressed, the trigger model continues to block 7
(the next block in the trigger model).
Also see
On event block
8-310
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 3-88)
Event constant
trigger.EVENT_NONE
trigger.EVENT_DISPLAY
trigger.EVENT_NOTIFYN
trigger.EVENT_COMMAND
trigger.EVENT_DIGION
trigger.EVENT_TSPLINKN
trigger.EVENT_LANN
trigger.EVENT_BLENDERN
trigger.EVENT_TIMERN
trigger.EVENT_ANALOGTRIGGER
trigger.EVENT_EXTERNAL
DMM7510-901-01 Rev. B / May 2015