Section 8: TSP command reference
Also see
Analog triggering overview
dmm.digitize.analogtrigger.pulse.polarity
dmm.measure.analogtrigger.mode
dmm.measure.analogtrigger.pulse.condition
dmm.measure.analogtrigger.pulse.polarity
dmm.measure.analogtrigger.pulse.width
dmm.measure.analogtrigger.pulse.width
This attribute defines the threshold value for the pulse width.
Type
TSP-Link accessible
Attribute (RW) Yes
Usage
value = dmm.measure.analogtrigger.pulse.width
dmm.measure.analogtrigger.pulse.width = value
value
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
This option is only available when the analog trigger mode is set to pulse.
This option sets either the minimum or maximum pulse width that generates an analog trigger event.
The value of pulse condition determines whether this value is interpreted as the minimum or
maximum pulse width.
8-126
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 3-64)
(on page 8-78)
(on page 8-121)
(on page 8-122)
(on page 8-125)
(on page 8-126)
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
The threshold value for the pulse width: 1 µs to 40 ms
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
Default value
1 ms
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
DMM7510-901-01 Rev. B / May 2015