[:Sense[1]]:<Function>:Azero[:State] - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
[:SENSe[1]]:<function>:AZERo[:STATe]
This command enables or disables automatic updates to the internal reference measurements (autozero) of the
instrument.
Type
Command and query
Usage
[:SENSe[1]]:<function>:AZERo[:STATe] <state>
[:SENSe[1]]:<function>:AZERo[:STATe]?
<function>
<state>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
To ensure the accuracy of readings, the instrument must periodically get new measurements of its
internal ground and voltage reference. The time interval between updates to these reference
measurements is determined by the integration aperture that is being used for measurements. The
Model DMM7510 uses separate reference and zero measurements for each aperture.
By default, the instrument automatically checks these reference measurements whenever a signal
measurement is made.
The time to make the reference measurements is in addition to the normal measurement time. If
timing is critical, you can disable autozero to avoid this time penalty.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the
drift, you can send the once command to make a reference and zero measurement immediately
before a test sequence.
For AC voltage and AC current measurements where the detector bandwidth is set to 3 Hz or 30 Hz,
autozero is set on and cannot be changed.
Example
VOLT:AZER OFF
Also see
Automatic reference measurements
[:SENSe[1]]:AZERo:ONCE
6-72
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
The status of autozero:
Disable autozero: OFF or 0
Enable autozero: ON or 1
RESistance
FRESistance
DIODe
CAPacitance
(on page 2-149)
(on page 6-114)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Sets autozero off for voltage measurements.
DMM7510-901-01 Rev. B / May 2015
Default value
ON (1)
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent

Advertisement

Table of Contents
loading

Table of Contents