[:Sense[1]]:<Function>:Relative:state - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
[:SENSe[1]]:<function>:RELative:STATe
This command enables or disables the application of a relative offset value to the measurement.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RELative:STATe <state>
[:SENSe[1]]:<function>:RELative:STATe?
<function>
<state>
Functions
VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
Details
When relative measurements are enabled, all subsequent measured readings are offset by the
relative offset value. You can enter a relative offset value or have the instrument acquire a relative
offset value.
Each returned measured relative reading is the result of the following calculation:
Displayed reading = Actual measured reading - Relative offset value
Example
:SENS:FUNC "VOLT"
:SENS:VOLT:REL 5
:SENSe:VOLT:REL:STATe ON
Also see
Relative offset
[:SENSe[1]]:<function>:RELative
[:SENSe[1]]:<function>:RELative:ACQuire
6-96
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which the setting applies; see Functions
Disable the relative offset: OFF or 0
Enable the relative offset: ON or 1
RESistance
FRESistance
DIODe
CAPacitance
(on page 3-4)
(on page 6-92)
Where saved
Save settings
Measure configuration list
TEMPerature
CONTinuity
FREQuency[:VOLTage]
PERiod[:VOLTage]
Set the measurement function to volts with a
relative offset of 5 V and enable the relative
offset function.
(on page 6-94)
Default value
0 (OFF)
VOLTage[:DC]:RATio
DIGitize:VOLTage
DIGitize:CURRent
DMM7510-901-01 Rev. B / May 2015