Keithley DMM7510 Reference Manual page 911

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference
Trigger events
Event description
Trigger event blender N (1 to 2), which combines trigger
events
Trigger timer N (1 to 4) expired
Analog trigger
External in trigger
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist() command to view the trigger model blocks in a list format.
The following usage has been deprecated; replace it with the usage above that includes the clear
parameter.
trigger.model.load("LoopUntilEvent", triggerEvent, position)
trigger.model.load("LoopUntilEvent", triggerEvent, position, delay)
trigger.model.load("LoopUntilEvent", triggerEvent, position, delay, bufferName)
trigger.model.load("LoopUntilEvent", triggerEvent, position, delay, bufferName,
readingBlock)
Example
reset()
-- Set up measure function
dmm.measure.func = dmm.FUNC_DC_CURRENT
-- Initiate readings
trigger.model.load("LoopUntilEvent", trigger.EVENT_DISPLAY, 50)
trigger.model.initiate()
Reset the instrument.
Set the instrument to measure current.
Load the LoopUntilEvent trigger model to make measurements until the front panel trigger key is pressed,
then continue to make measurements equal to 50 % of the reading buffer size.
Start the trigger model.
Also see
None
8-298
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Event constant
trigger.EVENT_BLENDERN
trigger.EVENT_TIMERN
trigger.EVENT_ANALOGTRIGGER
trigger.EVENT_EXTERNAL
DMM7510-901-01 Rev. B / May 2015

Advertisement

Table of Contents
loading

Table of Contents