Trigger:block:config:recall - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Example
TRIG:BLOC:CONF:PREV 14, "SETTINGS_LIST"
Also see
Configuration lists

:TRIGger:BLOCk:CONFig:RECall

This command recalls the system settings that are stored in a configuration list.
Type
Command only
Usage
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>"
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, "<configurationList>", <index>
<blockNumber>
<configurationList>
<index>
Details
When the trigger model reaches a configuration recall building block, the settings in the specified
configuration list are recalled.
You can restore a specific set of configuration settings in the configuration list by defining the index.
Example
TRIG:BLOCK:CONF:RECALL 1, "SETTINGS_LIST", 1
Also see
Configuration lists
DMM7510-901-01 Rev. B / May 2015
(on page 3-37)
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
A string that defines the configuration list to recall
The index in the configuration list to recall; defaults to 1 if not selected
(on page 3-37)
Section 6: SCPI command reference
Set trigger block 14 to restore the settings
from the previous index that is stored in the
configuration list SETTINGS_LIST.
Where saved
Save settings
Recall the settings in index 1 of the
SETTINGS_LIST configuration list as block
1 of the trigger model.
Default value
Not applicable
6-193