Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

dmm.measure.limit[Y].enable

This attribute enables or disables a limit test on the measurement from the selected measure function.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
state = dmm.measure.limit[Y].enable
dmm.measure.limit[Y].enable = state
state
Y
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
This command enables or disables a limit test for the selected measurement function. When this
attribute is enabled, the limit Y testing occurs on each measurement made by the instrument. Limit Y
testing compares the measurements to the high and low limit values. If a measurement falls outside
these limits, the test fails.
Example
See
dmm.measure.limit[Y].low.value
Also see
dmm.digitize.limit[Y].enable
dmm.measure.limit[Y].autoclear
dmm.measure.limit[Y].clear()
dmm.measure.limit[Y].fail
dmm.measure.limit[Y].high.value
dmm.measure.limit[Y].low.value
DMM7510-901-01 Rev. B / May 2015
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
Limit Y testing:
Disable: dmm.OFF
Enable: dmm.ON
Limit number: 1 or 2
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
(on page 8-164) for an example of how to use this command.
(on page 8-95)
(on page 8-159)
(on page 8-160)
(on page 8-162)
(on page 8-163)
(on page 8-164)
Section 8: TSP command reference
Default value
dmm.OFF
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE