Keithley DMM7510 Reference Manual page 557

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 6: SCPI command reference
Trigger events
Event description
Front-panel TRIGGER key press
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line <n> (1 to 6)
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
Appropriate LXI trigger packet is received on LAN trigger object
<n> (1 to 8)
Trigger event blender <n> (1 or 2), which combines trigger events BLENder<n>
Trigger timer <n> (1 to 4) expired
Analog trigger
External in trigger
Example
*RST
SENS:FUNC "CURR"
TRIG:LOAD "LoopUntilEvent", DISP, 25
INIT
Also see
None
6-230
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Reset the instrument.
Set the instrument to measure DC current.
Set the LoopUntilEvent trigger model to make
measurements until the front-panel TRIGGER
key is pressed after starting the trigger model,
then make measurements that constitute 75 % of
the reading buffer.
Start the trigger model.
Event constant
DISPlay
NOTify<n>
COMMand
DIGio<n>
TSPLink<n>
LAN<n>
TIMer<n>
ATRigger
EXTernal
DMM7510-901-01 Rev. B / May 2015

Advertisement

Table of Contents
loading

Table of Contents