Section 8: TSP command reference

trigger.timer[N].start.seconds

This attribute configures the seconds of an alarm or a time in the future when the timer will start.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
time = trigger.timer[N].start.seconds
trigger.timer[N].start.seconds = time
time
N
Details
This command configures the alarm of the timer.
When the timer is enabled, the timer starts immediately if the timer is configured for a start time that
has passed.
Example
trigger.timer[1].start.seconds = localnode.gettime() + 30
trigger.timer[1].enable = trigger.ON
Set the trigger timer to start 30 s from the time when the timer is enabled.
Also see
None

trigger.timer[N].start.stimulus

This attribute describes the event that starts the trigger timer.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
event = trigger.timer[N].start.stimulus
trigger.timer[N].start.stimulus = event
event
N
8-336
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
Trigger timer N reset
The time: 0 to 2,147,483,647 s
Trigger timer number (1 to 4)
Affected by
Restore configuration
Instrument reset
Power cycle
Trigger timer N reset
The event that starts the trigger timer
Trigger timer number (1 to 4)
Where saved
Default value
Configuration script
0
Where saved
Default value
Configuration script
trigger.EVENT_NONE