Section 2: General operation
Figure 95: Rear panel connections: Digitize current measurement (current below 3 A)
Figure 96: Rear panel connections: Digitize current measurement (current below 10 A)
Measure with digitize current using the front panel
To make a digitize current measurement using the front panel:
1. Make the connections as shown in the previous figures.
2. Press the FUNCTION key.
3. Select the Digitize Functions tab.
4. Select Digitize Current.
5. Press the TRIGGER key for 2 seconds and verify that the instrument is set to Continuous
Measurement.
The measurements start displaying on the front panel.
Settings available for digitize current measurements
See
Digitize current measure settings
digitizing current measurements.
2-130
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
(on page 2-33) for settings that are available when you are