Dmm.measure.threshold.range - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Section 8: TSP command reference

dmm.measure.threshold.range

This attribute indicates the expected input level of the voltage signal.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
value = dmm.measure.threshold.range
dmm.measure.threshold.range = value
value
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_DC_CURRENT
dmm.FUNC_AC_CURRENT
dmm.FUNC_TEMPERATURE
Details
The range setting conditions the signal. The instrument automatically selects the most sensitive
threshold range for the value you enter. For example, if you specify the expected input voltage to be
90 mV, the instrument automatically selects the 100 mV threshold range.
Example
dmm.measure.threshold.range = 50
Also see
dmm.measure.threshold.autorange
dmm.measure.threshold.level
8-196
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
The range: 0.1 to 700; instrument selects nearest valid range (100 mV, 1 V, 10 V,
100 V, 700 V)
dmm.FUNC_RESISTANCE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_DIODE
dmm.FUNC_CAPACITANCE
dmm.FUNC_CONTINUITY
(on page 8-194)
(on page 8-195)
Default value
10 V
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DCV_RATIO
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Set the threshold range for the selected
function to 100 V.