Display:reading:format - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Details
This command changes the light output of the front panel when a test requires different instrument
illumination levels.
The change in illumination is temporary. The normal backlight settings are restored after a power
cycle. You can use this to reset a display that is already dimmed by the front-panel Backlight Dimmer.
Screen life is affected by how long the screen is on at full brightness. The higher the brightness
setting and the longer the screen is bright, the shorter the screen life.
Example
DISP:LIGH:STAT ON50
Also see
Adjust the backlight brightness and dimmer

:DISPlay:READing:FORMat

This command determines the format that is used to display measurement readings on the front-panel display of
the instrument.
Type
Command and query
Usage
:DISPlay:READing:FORMat <format>
:DISPlay:READing:FORMat?
<format>
Details
This setting persists through *RST and power cycles.
When Prefix is selected, prefixes are added to the units symbol, such as k (kilo) or m (milli). When
Exponent is selected, exponents are used instead of prefixes. When the prefix option is selected, very
large or very small numbers may be displayed with exponents.
Example
DISP:READ:FORM EXP
Also see
Setting the display format
DMM7510-901-01 Rev. B / May 2015
(on page 2-10)
Affected by
Not applicable
Use exponent format: EXPonent
Add a prefix to the units symbol, such as k, m, or µ: PREFix
(on page 2-56)
Section 6: SCPI command reference
Set the display brightness to 50 %.
Where saved
Nonvolatile memory
Change front-panel display to show readings in
exponential format.
Default value
PREF
6-45

Advertisement

Table of Contents
loading

Table of Contents