Trigger:load "Keithley2001 - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

:TRIGger:LOAD "Keithley2001"

This command loads a predefined trigger model configuration that emulates a Keithley Instruments 2001 trigger
model.
Type
Command only
Usage
:TRIGger:LOAD
<arm2Bypass>, <arm2Source>, <arm2Count>, <arm2Delay>, <trigBypass>,
<trigSource>, <trigCount>, <trigDelay>
<arm1Bypass>
<arm1Source>
<arm1Count>
<arm2Bypass>
<arm2Source>
<arm2Count>
<arm2Delay>
<trigBypass>
<trigSource>
<trigCount>
<trigDelay>
DMM7510-901-01 Rev. B / May 2015
Affected by
Not applicable
"Keithley2001", <arm1Bypass>, <arm1Source>, <arm1Count>,
Bypass Arm 1: ON (1)
Do not bypass Arm 1: OFF (0)
The event that triggers Arm 1; see Details
The number of times to repeat Arm 1
Bypass Arm 2: ON
Do not bypass Arm 2: OFF
The event that triggers Arm 2; see Details
The number of times of times to repeat Arm 2
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
Bypass the trigger layer: ON
Do not bypass the trigger layer: OFF
The event that triggers the trigger layer; see Details
The number of times to repeat the trigger layer
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
Section 6: SCPI command reference
Where saved
Not applicable
Default value
Not applicable
6-225