Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
The instrument makes two sets of readings. The first set is made until the trigger event occurs. The
second set is made after the trigger event occurs, up to the number of readings calculated by the
position parameter.
You cannot have the event constant set at none when you run this predefined trigger model.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
Trigger events
Event description
Front-panel TRIGGER key press
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
(1 to 6)
Line edge detected on TSP-Link synchronization line N
(1 to 3)
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)
DMM7510-901-01 Rev. B / May 2015
Section 8: TSP command reference
Event constant
trigger.EVENT_DISPLAY
trigger.EVENT_NOTIFYN
trigger.EVENT_COMMAND
trigger.EVENT_DIGION
trigger.EVENT_TSPLINKN
trigger.EVENT_LANN
8-297