Trigger:load "Logictrigger - Keithley DMM7510 Reference Manual

7 1/2 digit graphical sampling multimeter
Hide thumbs Also See for DMM7510:
Table of Contents

Advertisement

Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual

:TRIGger:LOAD "LogicTrigger"

This command loads a predefined trigger model configuration that sets up an external trigger through the digital
I/O.
Type
Command only
Usage
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <delay>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <delay>,
"<bufferName>"
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <delay>,
"<bufferName>", <readingBlock>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>, <delay>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>, <delay>,
"<bufferName>"
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>, <delay>,
"<bufferName>", <readingBlock>
<digInLine>
<digOutLine>
<count>
<clear>
<delay>
<bufferName>
<readingBlock>
Details
This trigger model waits for a digital input or external trigger input event to occur, makes a
measurement, and issues a notify event. If a digital output line is selected, a notify event asserts a
digital output line. A notify event asserts the external output line regardless of the line settings. You
can set the line to 7 to assert only the external output line, or to another setting to assert both a digital
output line and the external output line.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
TRIGger:BLOCk:LIST? command to view the trigger model blocks in a list format.
This command replaces the :TRIGger:LOAD "ExtDigTrigger" command, which is deprecated.
DMM7510-901-01 Rev. B / May 2015
Affected by
Not applicable
The digital input line (1 to 6) or external input line (7); also the event that
the trigger model will wait on in block 1
The digital output line (1 to 6) or external input line (7)
The number of measurements the instrument will make
To clear previously detected trigger events when entering the wait block:
ENTer
To immediately act on any previously detected triggers and not clear them
(default): NEVer
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
The name of the reading buffer, which may be a default buffer
(defbuffer1 or defbuffer2) or a user-defined buffer; default is
defbuffer1
Define a measure or digitize block for the trigger model; options are:
ACTive: Add a measure or digitize block to the trigger model based
on the active function; if no option defined, ACTive is used
MEASure: Adds a measure block to the trigger model
DIGitize: Adds a digitize block to the trigger model