Left Rear Wire Mesh Filter - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Fans and Filters

Left Rear Wire Mesh Filter

406
TSQ 9610 Hardware Manual
To replace the left rear wire mesh filter
Note See the ISQ and TSQ GC-MS Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
on
page
139.
Note You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T10 Torxhead screwdriver to remove the four screws holding the left rear wire mesh
filter to the left hand sub panel.
Figure 445. Replacing the Left Rear Wire Mesh Filter
3. Remove and replace the left rear wire mesh filter.
4. Reattach the four screws holding the left rear wire mesh filter to the left hand sub panel.
5. Restart your system by following the instructions in
"Shutting Down the TSQ 9610 System"
Screws
"Restarting the System"
on
page
158.
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents