Calibration Gas Module

Thermo Scientific
The EI calibration gas module requires little maintenance, except to refill the calibrant in the
reservoir. Your calibration gas controller may need to be replaced if you add too much
calibrant and you contaminate the system. The process is the same for replacing a single- or
dual-flow calibration gas module.
To replace the EI calibration gas module
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139.
Note You do not have to remove the ion source cartridge or column.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 418. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
3
Replacing the Calibration Gas Components
"Shutting Down the TSQ 9610 System"
Screws
TSQ 9610 Hardware Manual
Advanced Troubleshooting
Screw on Front
of Instrument
381