Advanced Troubleshooting
Replacing Components of the Ion Source
208
TSQ 9610 Hardware Manual
10. Reconnect the lens heater block grounding strap to the lens heater block
Figure 208. Reconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
11. Reconnect the Q1 entrance lens, Q1 exit lens, Q3 entrance lens, and Q3 exit lens to the
source interface board using a flat-tipped screwdriver.
12. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 9610 Instrument"
instrument.
13. Restart your system by following the instructions in
on
page 154
to replace the tray in the TSQ
"Restarting the System"
"Replacing the
on
page
158.