Thermo Scientific
6. Hold the new source interface board next to the old board.
7. As you remove each of the wires from the old board, connect them to the same connectors
on the new board.
Figure 405. Reconnecting the Source Interface Board Wires
8. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray in the TSQ 9610 Instrument"
tray in the TSQ instrument.
9. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
on
page 154
to replace the
"Restarting the System"
TSQ 9610 Hardware Manual
on
page
158.